Skip to content

Public FDA record

Z-1681-2022: Product Description/Item: ChemoLock Bag Spike, 10 units/CL-10-10; ChemoLock BAG SPIKE w/ ADDITIVE PORT, D

Recall record Z-1681-2022 for ICU Medical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1681-2022
Company in FDA record
ICU Medical, Inc.
FDA record creation date
Sep 7, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Ports have a potential for unintended disconnection, inability to connect to or fully engage with injectors, due to a variation in the spring inside the Port, which may cause delay of therapy and exposure to caustic substances. Ports available as standalone connectors, or utilized as access points on vial adaptors, bag spikes, and administration sets.

Source facts

Action
On 06/27/2022 recall notices were mailed to customers who were asked to do the following: 1) Discontinue the use and distribution of the affected product immediately. Check your inventory and quarantine all affected product at your facility. 2) In instances where the benefits of using the ChemoLock are greater than the potential risks for non-connection or disconnection, and you choose to utilize ChemoLock affected lot numbers, ensure full engagement between the ChemoLock Port and ChemoLock Injector. If full engagement is achieved, the device will work as intended. If full engagement cannot be achieved or is difficult to achieve, discard product and utilize a new device. 3) Complete and return the response form. 4) The recalling firm's customer service can be contacted for product availability information. 5) Return affected product using the return label provided. Contact Sedgwick if you have not received a return label or require additional labels. 6) If you have distributed the product further, immediately notify your accounts and ask them to contact Sedgwick. For further inquiries, please contact: Global Complaint Management 1-844-654-7780 or ProductComplaintsPP@icumed.com; Customer Service 1-866-829-9025, option 8 or customerservice@icumed.com; Sedgwick 1-888-943-5190 (M-F, 8am-5pm ET)
Initiated
2022-06-27
Root cause
Component design/selection
Distribution
Worldwide distribution - US Nationwide distribution including in the states of LA, FL, TX, NY, NE, WI, PA, IA, ND, CA, OH, WA, GA, AL, CO, AZ, NC, TN, MI, VA, MA, DE, OR, KY, CT, OK, NJ, MN, IL, VT, AR, RI, ID, IN, UT, NV, MT, MD, KS, SC, NM, ME, MO, SD and the countries of : Saudi Arabia, Australia, United Arab Emirates, Spain, Slovenia, Japan, Canada.
Recall number
Z-1681-2022
Recall status
Open, Classified
CDRH record ID
195036
FDA record created
2022-09-07
Recall event number
90606

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.