Z-1682-2015: Ingenuity TF Positron Emission Tomography (PET) and Magnetic Resonance Imaging (MRI), 882380, Philips Med

Recall record Z-1682-2015 for Philips Medical Systems (Cleveland) Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1682-2015
Company in FDA record
Philips Medical Systems (Cleveland) Inc
FDA classification date
May 28, 2015
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

It was noticed that the spectroscopy voxel map did not align with the anatomical images of the phantom. This resulted in incorrect orientation of the Chemical Shift Imaging (CSI) data.

Source facts

Action
On 3/11/2015 the firm sent Urgent - Medical Device Correction letters to their customers. The letters identified the affected device, explained the reason for the recall, how to identify affected product, advice on actions to be taken by customer/user, and any actions planned by the firm. Customers were advised to use software provided by Philips to view spectroscopy images; and, not to export and view images using third party software. The firm stated that it will be issuing a software update. If further information or support concerning the issue is needed, customers are asked to contact their local Philips representative or local Philips Healthcare office.
Initiated
2015-03-11
Root cause
Software in the Use Environment
Terminated
2018-06-26
Distribution
Worldwide Distribution -- US, including the states of NY, OH, & TX; and, the countries of Finland, Germany, Japan, Netherlands, South Korea, Spain, & Switzerland.
Recall number
Z-1682-2015
Recall status
Terminated
CDRH record ID
136871
FDA classified
2015-05-28
Recall event number
71228

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗