Open original FDA record ↗Recall record
- FDA identifier
- Z-1682-2017
- Company in FDA record
- Hospira Inc., A Pfizer Company
- FDA record creation date
- Mar 24, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Depleted Battery alarm shows instead of Replace Battery. On battery power, ongoing therapy stops, alarms show and sound, pump shuts down after 3 minutes. On AC power, the pump reboots (takes about 25 seconds). Therapy can resume. AC power interruption of 3-7 seconds prompts an incorrect E323 alarm, ongoing therapy stop, and pump reboot. Longer power loss results in battery power operation.
Source facts
- Action
- On 12/30/2016, firm sent notification of the recall to Hospira direct consignees of record via traceable mailing. Delivery will be confirmed via traceable mailing. Consignees who further distributed the product were asked to notify their customers. Effectiveness of the communication will be confirmed by phone call with direct consignees who do not respond. The final disposition of the recalled product will be field correction (firm will contact consignees to upgrade the software in their Plum 360 infusion pumps).
- Initiated
- 2016-12-30
- Root cause
- Software in the Use Environment
- Terminated
- 2019-07-09
- Distribution
- Nationwide, Australia, New Zealand, Canada, Mexico, Chile, Philippines, Hong Kong, Greece, Ireland, Jordan, Kuwait, Oman, Poland, Saudi Arabia, Turkey, United Arab Emirates, Italy, France, United Kingdom, Spain
- Recall number
- Z-1682-2017
- Recall status
- Terminated
- CDRH record ID
- 152590
- FDA record created
- 2017-03-24
- Recall event number
- 76253