Z-1683-2023: QIAstat-Dx Respiratory SARS-CoV-2 Panel multiplexed nucleic acid real-time PCR test, Reference Number 691

Recall record Z-1683-2023 for Qiagen Sciences LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1683-2023
Company in FDA record
Qiagen Sciences LLC
FDA classification date
Jun 2, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Faulty cartridges in the lot could result in false test results.

Source facts

Action
An Urgent Medical Device Correction notification letter dated April 2023 was sent to customers. Actions to be taken by customer/user Please check if you have remaining stock of LOT 175011350. Please DO NOT use it. Dispose of it immediately in accordance to your national and local safety and environmental regulations. Please contact QIAGEN Technical Services for a free-of-charge replacement. If you already used cartridges from this LOT, please identify the results obtained with cartridges from the affected SN. For results obtained with an affected SN, review the results as follows: o Review all results of the respective targets mentioned in this notice to exclude erroneous diagnosis and treatment, except in those cases where alternative confirmation was obtained. o Review any result for which tests were performed on clinical and/or epidemiological suspicion of the respective targets. Review this notice with your laboratory/medical director. IMPORTANT: Forward this information to all individuals and departments within your organization using the above listed kits. If you are not the end user, please forward this notice to the product end user. Complete the Acknowledgement of Receipt Form attached to this letter by 14 May 2023. Actions taken by QIAGEN QIAGEN was able to identify the root cause of the issue and has implemented immediate actions to ensure that the currently produced cartridges will not cause this issue moving forward. If you have any questions or concerns, please contact your local QIAGEN Technical Services Department through any of the following: Telephone: 800 362 7737 Email: TechService-NA@qiagen.com
Initiated
2023-04-28
Root cause
Under Investigation by firm
Distribution
US Nationwide distribution in the states of CA, FL, IL, MD, MS, NY, OK, PA, TN.
Recall number
Z-1683-2023
Recall status
Open, Classified
CDRH record ID
200158
FDA classified
2023-06-02
Recall event number
92298

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗