Open original FDA record ↗Recall record
- FDA identifier
- Z-1684-2017
- Company in FDA record
- Streck
- FDA classification date
- Mar 8, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Complaints of the 2.0ml tube stopper coming loose during phlebotomy when small diameter tube adapters were not utilized. Such adapters aid with centering small diameter tubes to properly pierce the middle of the stoppers.
Source facts
- Action
- On February 23, 2017, Streck electronically notified customers of the updated device correction. The recalling firm used an electronic mail alert to notify all of their customers the IFU for the recalled product had been revised. The specified the revision included phlebotomy guidelines specific to the 2.0 ml blood collections tubes. The revised IFU was provided as a link in the email. The recalling firm is reportedly able to monitor the access of the revised IFU to ensure each customer has received the information. For further questions, please call (402) 333-1982.
- Initiated
- 2017-02-23
- Root cause
- Error in labeling
- Distribution
- Worldwide Distribution - USA (nationwide) and to the countries of : Belgium, Canada, Germany, Israel, Japan, the Republic of Korea, the Netherlands, and the United Kingdom. There was no military or government distribution.
- Recall number
- Z-1684-2017
- Recall status
- Open, Classified
- CDRH record ID
- 153381
- FDA classified
- 2017-03-08
- Recall event number
- 76501