Z-1684-2020: ABL800 FLEX, model number/UDI 393-800/0570 In Vitro testing of samples of whole blood for the parameters

Recall record Z-1684-2020 for Radiometer Medical ApS. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1684-2020
Company in FDA record
Radiometer Medical ApS
FDA record creation date
Apr 8, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Analyzer's barcode reader misinterprets the contents of barcode label used for entering patient ID or accession number. The issue is related to barcode types not using a check digit. This could result in patient mixup or loss of sample resulting in delayed medical treatment

Source facts

Action
On 3/10/2020, a "Urgent: medical Device Recall" letter was sent via email and FedEx to affected consignees. In addition to informing the consignee about the recall issue, the recall notification asked customers to take the following action: Check if your institution is using barcode types without a check digit, including e.g. patient ID or accession number, to be read on the ABL800. 1. If you use barcode types without a check digit, Radiometer recommends to either: - Enable the check digit on the barcode type currently used, or - Change type of barcode to one that includes a check digit, as this enables the barcode reader to validate the data read, and hence, to capture if the barcode has been misinterpreted and consequently to reject such data. 2. If you use barcode types with a check digit the barcode reader already validates the data read, and hence, captures if the barcode has been misinterpreted and consequently rejects such data. Hence, no action required. 3. If you do not use barcodes to be read by the ABL800, no action required. 4. Complete the Recall Response Form (last page of this letter) and submit to your Radiometer distributor. 5. If you have any questions, please contact us at 1-800-736-0600 option 1, Monday Friday 9am EST to 7pm EST. Solution provided by Radiometer Radiometer has released software version V6.19, which we will install on your analyzer, disregarding which barcode type is used, or no barcodes are used at all. The new software version scans the barcode three times (as opposed to once in previous versions). For barcode types without a check digit, this change may potentially reduce the risk of misinterpretation, if caused by the factors, which may add to the risk of misinterpretation, mentioned above.
Initiated
2020-03-10
Root cause
Software design
Terminated
2022-10-31
Distribution
Worldwide distribution - US Nationwide distribution in the states of AL, AR, AZ, BM, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY and countries of Canada, China, Czech Republic, Denmark, Germany, Spain, France, Hungary, United Kingdom, Australia, Poland, South Africa, Singapore, Switzerland, Turkey, Norway, Austria, Italy, Sweden, Finland, Croatia, Russia, Norway.
Recall number
Z-1684-2020
Recall status
Terminated
CDRH record ID
180210
FDA record created
2020-04-08
Recall event number
85168

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗