Z-1684-2022: Early Bird Bleed Monitoring System Introducer Sheath Set, REF: FG001-01 (6F), FG001-02 (8F)

Recall record Z-1684-2022 for Saranas, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1684-2022
Company in FDA record
Saranas, Inc.
FDA record creation date
Sep 7, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Bleed monitoring system device design error, causing the device to not power on and complete the series of self-tests so instructions for use being updated to include device check before use ensure that the device is functioning appropriately.

Source facts

Action
On 8/10/22 correction notices were emailed to customers who were asked to do the following: 1. Review the updated Instructions for Use, specifically, "12 DIRECTIONS FOR USE" 2. Prior to the operation as part of device preparation, activate the device by pulling the battery isolation pull tab and observing that the green Power indicator is blinking, and no other lights remain on. Once confirmed, immediately press and hold the Power button to turn off the device. 3. During the power-on sequence, if there is an error, the device will sound an audible tone and the yellow error light will remain illuminated. The user should attempt to restart the device by pressing the Power button to turn off the device and turn the device back on for one additional attempt for use of the device. If the device does not power back on as intended (blinking green light only), the user must discontinue use of the device. If the device is defective, the user should obtain a new device, if available. The same safety check should be performed on the new device. The defective device should be returned to the recalling firm. 4. Complete and return the recall return response form Customers with additional questions are encouraged to contact the recalling firm at info@saranas.com or 1-833-ERLYBRD (833-375-9273).
Initiated
2022-08-10
Root cause
Nonconforming Material/Component
Distribution
US Nationwide distribution including in the states of PA, MI, FL, WV, TX, NJ, GA, LA, NY, MA, CT, AZ, OH, TN, VA.
Recall number
Z-1684-2022
Recall status
Open, Classified
CDRH record ID
195571
FDA record created
2022-09-07
Recall event number
90725

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗