Open original FDA record ↗Recall record
- FDA identifier
- Z-1685-2012
- Company in FDA record
- Siemens Medical Solutions USA, Inc.
- FDA classification date
- May 30, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
When using the Advanced SieClear spatial compounding feature a synchronization error may occur which can cause a mismatch between the image and the depth scale.
Source facts
- Action
- Siemens initiated a Customer Safety Advisory Notification letter Field correction on April 23, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers' local Customer Service Engineer will install an update to their system as soon as it becomes available. For questions customers were instructed to contact their local service support person. Customers were also asked to pass the notification leter to all those within their organization who need to be aware of the issue until the corrective action is completed. For questions regarding this recall call 650-694-5747.
- Initiated
- 2012-04-23
- Root cause
- Software change control
- Terminated
- 2013-07-16
- Distribution
- Worldwide Distribution - USA (nationwide) and the countries of Argentina, Austria, Australia, Belgium, Canada, Switzerland, Chile, China, Germany, Denmark, Ecuador, Spain, France, Great Britain, Greece, Hong Kong, Hungary, Indonesia, Israel, India, Italy, Jordan, Japan, South Korea, Kuwait, Lebanon, Myanmar, Mexico, Malaysia, Netherlands, Norway, New Zealand, Peru, Philippines, Poland, Portugal, Romania, Russian Federation, Saudi Arabia, Sweden, Slovenia, El Salvador, and Thailand.
- Recall number
- Z-1685-2012
- Recall status
- Terminated
- CDRH record ID
- 109143
- FDA classified
- 2012-05-30
- Recall event number
- 61808