Open original FDA record ↗Recall record
- FDA identifier
- Z-1685-2025
- Company in FDA record
- Siemens Healthcare Diagnostics Inc
- FDA classification date
- Apr 29, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The occurrence of discrepant high pH results in samples introduced with higher injection volumes with epoc sensor configuration 45.n.
Source facts
- Action
- Consignees were sent an Urgent Medical Device Correction notice dated March 2025. The notification instructs consignees to review the provided notice with their medical director to determine the appropriate course of action. If affected devices were further distributed consignees are asked to forward the recall notification. Consignees are to return the provided Field Correction Effectiveness Check form within 7 days of receipt. The firm recommends that consignees stop injecting the sample as soon as the audio/visual cues are presented to minimize the potential for erroneous pH results. The current affected software expires on 06/10/25 and the upgrade will include a correction for this issue. Consignees are to retain the recall notification with their manual or near the system for reference. Consignees with questions can call the Siemens Healthineers Technical Solutions Center at 877-229-3711.
- Initiated
- 2025-03-19
- Root cause
- Process control
- Distribution
- Domestic: Nationwide Distribution; Foreign: Albania, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bolivia, Bosnia, Brazil, Cambodia, Canada, Canary Islands, Chile, China, Colombia, Croatia, Ecuador, Estonia, France, Germany, Ghana, Greece, Guadeloupe, Guatemala, Hong Kong, Hungary, Iceland, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kuwait, Lithuania, Malaysia, Mali, Martinique, Mexico, Morocco, Netherlands, New Caledonia, New Zealand, Nigeria, Oman, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Federation, San Marino, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Vietnam.
- Recall number
- Z-1685-2025
- Recall status
- Open, Classified
- CDRH record ID
- 213125
- FDA classified
- 2025-04-29
- Recall event number
- 96539