Open original FDA record ↗Recall record
- FDA identifier
- Z-1686-2009
- Company in FDA record
- Depuy Orthopaedics, Inc.
- FDA classification date
- Aug 21, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
There may be a crack on the lateral side of the condyle in the posterior chamfer region. The device is used as a Orthopedic knee implant.
Source facts
- Action
- The firm's sales staff was notified by email on May 7, 2009, and the recall was expanded to include additional lots on May 29, 2009, instructing them to remove the affected lots from their inventory and the inventory of any consignees, to provide the consignees with a copy of the recall letter, and to return the product to the firm. Hospital consignees were notified by letter dated May 7, 2009, as well as by updated letter on May 29, 2009. A dear doctor letter dated May 29, 2009 was issued to implanting surgeons advising them of the issue. Direct questions to Depuy Customer Service by calling 1-800-366-8143.
- Initiated
- 2009-05-07
- Root cause
- Process control
- Terminated
- 2010-10-25
- Distribution
- Nationwide.
- Recall number
- Z-1686-2009
- Recall status
- Terminated
- CDRH record ID
- 82075
- FDA classified
- 2009-08-21
- Recall event number
- 52088