Skip to content

Public FDA record

Z-1686-2010: Juxta-Lite compression wrap XL Short 28cm Length with Anklet Catalog 23026017

Recall record Z-1686-2010 for Circaid Medical Products Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1686-2010
Company in FDA record
Circaid Medical Products Inc
FDA classification date
May 27, 2010
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Some systems in these lots have Built-In Pressure System lines with the incorrect amount of spacing. Some systems in these lots had I inch space between each line with no tension applied. The system should have 1.25 inches of space. The correct spacing (1.25") is necessary to achieve the measured tension ranges. The incorrect spacing may result in the inability to achieve the correct tension level

Source facts

Action
CircAid Medical Products Important Device Product Recall letter dated Apr 20, 2010 was mailed to consignees describing problem, listing affected product and distribution dates. Customers were instructed to locate and return these Juxta-Lite Systems from stock and notify their customers A Retum Authorization number and contact name was provided. this listing for product returns.
Initiated
2010-04-20
Root cause
Process control
Terminated
2011-01-06
Distribution
Worldwide distribution: USA, Japan, Canada, Netherlands
Recall number
Z-1686-2010
Recall status
Terminated
CDRH record ID
91051
FDA classified
2010-05-27
Recall event number
55529

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗