Open original FDA record ↗Recall record
- FDA identifier
- Z-1686-2015
- Company in FDA record
- Medrad Inc dba Bayer R&I
- FDA classification date
- May 28, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The root cause investigation showed that the under-volume hazard can occur when the unit has been configured to prime with contrast, but instead primes with saline due to the configuration data not being sent to the injector head. This may create a non-optimal priming condition that could lead to under-infusion.
Source facts
- Action
- Bayer Healthcare sent a Medical Device Field Corrective Action letter to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to identify any affected device. Bayer Service will schedule time to update their software free of charge. Customers with questions were instructed to contact the customer support team at 1-800-633-7237. For questions regarding this recall call 412-767-2400.
- Initiated
- 2013-09-20
- Root cause
- Software design
- Terminated
- 2015-06-16
- Distribution
- Nationwide Distribution
- Recall number
- Z-1686-2015
- Recall status
- Terminated
- CDRH record ID
- 137254
- FDA classified
- 2015-05-28
- Recall event number
- 71296