Skip to content

Public FDA record

Z-1686-2016: Diamedix Is-EBV-VCA IgG Test Kit Diamedix Is-Mumps IgG Test Kit is an EIA procedure intended for the qu

Recall record Z-1686-2016 for Diamedix Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1686-2016
Company in FDA record
Diamedix Corporation
FDA record creation date
May 16, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A gel-like substance was found in kit component, sample diluent , that could potentially compromise product performance.

Source facts

Action
The firm, Diamedix, sent letters dated October 14, October 20 and December 5, 2014 to its customers directly informing them of the correction action. Notification of all domestic customers via Certified Mail. Notification fo the international customers via Federal Express and/or e-mail. With the letter, a replacement bottle of Sample B Diluent was sent and the customer was advised to discard any unused bottles of affected product. New lot number labels were provided to identify kits with the replacement Sample B Diluent. Customers with questions were told they could contact Diamedix Technical Services (1-800-327-4565) if they had any questions or required any additional guidance regarding the correction.
Initiated
2014-10-20
Root cause
Unknown/Undetermined by firm
Terminated
2016-12-14
Distribution
Worldwide distribution: US nationwide including states of: AZ, CA, CO, FL, GA, IL, IN, KY, MA, MI, MO, NC, NE, NH, NJ, NY, OH, OK, OR, PA, RI, TN, TX, VA, WI and Puerto Rico. and Country of: ITALY.
Recall number
Z-1686-2016
Recall status
Terminated
CDRH record ID
144775
FDA record created
2016-05-16
Recall event number
73694

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗