Z-1687-2019: Abbott MitraClip XTR Clip Delivery System, UDI: 08717648226366, Part: CDS0601-XTR, Sterile EO, Rx Only

Recall record Z-1687-2019 for Abbott Vascular. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1687-2019
Company in FDA record
Abbott Vascular
FDA record creation date
May 29, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Reports of implantable mitral valve repair system clips unexpectedly opening and becoming nonfunctional, resulting from unintended excessive force applied during the Clip implantation. Excessive force can also result in unexpected movement of clip arms. The inability to close and remove the device has lead to surgery and additional intervention.

Source facts

Action
On 05/01/19, Urgent Medical Device Correction notices were sent to implanting physicians, via personal delivery or email, informing them that under normal use conditions, this failure will not occur. To prevent unintended force from being applied to the clip, revised instructions, including definition and related technique, for performing the steps "Establish Final Angle" and "Invert the Clip Arms" were developed and provided herein. The revised instructions will be applied to the shared XTR and NTR Clip Delivery System Instructions for Use. Unintended excessive force applied to the NTR Clip Delivery System during "Establish Final Arm Angle" can result in unexpected movement of clip arms; however, unlike XTR, the NTR Clip does not become damaged and remains functional. Adhering to the revised IFU sections for both product platforms will ensure consistency of use. Further, the firm requested that customers read the Field Safety Notice, review revised instructions for use with firm representatives, sign and return Effectiveness Check Forms, and share this information with all personnel associated with the procedure. The revised instruction steps provided in this communication will be updated in the product IFU and included in associated training. Your current inventory of product is acceptable for safe use following the revised IFU steps described above. There is no need to return any product. Customers with additional questions are encouraged to call customer service at 800-227-9902.
Initiated
2019-05-01
Root cause
Use error
Terminated
2020-12-11
Distribution
U.S.: FL, UT, IN, OK, TX, CO, LA, AZ, MO, AL, GA, CA, AR, KS, IL, HI, IA, KY, MA, DC, NY, NV, NM, NJ, MT, MD, NC, CT, MI, MN, OH, ND, NH, OR, PA, WI, WA, VA, WV, TN, SC, RI, DE, ME, NE, SD, MS, ID, VT. Foreign (OUS): AUSTRALIA, AUSTRIA, BELGIUM, BULGARIA, Canada, CROATIA, CZECH REPUBLIC, DENMARK, FINLAND, FRANCE, GERMANY, GREECE, HONG KONG, HUNGARY, INDONESIA, IRELAND, ISRAEL, ITALY, KUWAIT, LEBANON, MACAU, NETHERLANDS, NEW ZEALAND, NORWAY, POLAND, PORTUGAL, ROMANIA, RUSSIA, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SPAIN, SWEDEN, SWITZERLAND, THAILAND, TURKEY, UNITED KINGDOM
Recall number
Z-1687-2019
Recall status
Terminated
CDRH record ID
172622
FDA record created
2019-05-29
Recall event number
82746

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗