- Action
- On 5/1/2015 Terumo issued an URGENT MEDICAL DEVICE CORRECTION letter to users explaining that Terumo has identified causes of the Terumo System 1 experiencing loss of system power. These activities will be implemented in a phased approach beginning late 2015:
- Implementing design changes and upgrades to improve the reliability of power switching and battery backup.
- Adding a switch protector to the On/Off power switch to prevent inadvertent shut off.
- Replacing the On/Off power switch with a new design to prevent switch failure.
A Terumo Field Service Representative will contact users to schedule the field correction activities once the activities are ready to be implemented.
CUSTOMER INSTRUCTIONS: Review this Medical Device Correction and assure that all users have received notice of this issue.
- Complete and return the enclosed Customer Response Form.
- Terumo CVS will contact users to schedule the field correction activities.
We encourage you to contact us with any questions or concerns:
Terumo CVS Customer Service
1-800-521-2818
Fax 1-734-741-6149
Customer Service Hours:
Monday Friday, 8 a.m. 6 p.m. ET
- Initiated
- 2015-05-01
- Root cause
- Device Design
- Terminated
- 2018-09-28
- Distribution
- Worldwide Distribution-US (nationwide) including DC and the states of AL , AR , AZ , CA , CO , CT , DE , FL , GA , HI , IA , ID , IL , IN , KY , LA , MA , MD , ME , MI , MN , MO , MS , MT , NC , ND , NE , NJ , NM , NV , NY , OH , OK , OR , PA , SC , TN , TX , UT , VA , VT , WA , WI , and WV, and the countries of Mexico , AUSTRALIA , UNITED ARAB EMIRATES (UAE) , Hong Kong , Indonesia , Singapore , Taiwan , Thailand , Philippines , South Korea , CHILE , Vietnam , Malaysia , BELGIUM , Japan , and CANADA.
- Recall number
- Z-1688-2015
- Recall status
- Terminated
- CDRH record ID
- 137277
- FDA classified
- 2015-05-28
- Recall event number
- 67989