Z-1688-2020: "HEMOGLOBIN A1c CALIBRATORS *** REF 4P52-01 *** 307261/R03 *** S4P5X0 *** FOR USE WITH ARCHITECT" - Produ

Recall record Z-1688-2020 for Abbott Laboratories. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1688-2020
Company in FDA record
Abbott Laboratories
FDA record creation date
Apr 9, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Calibrator level 2 bottle may have been manufactured incorrectly for a portion of the lot. The manufacturer observed no impact to assay performance between the HbA1c concentration range of 5.33% to 6.87%. However, outside of this range, a bias of greater than 3% may be seen with patient results. The device failure that is associated with this recall can result in potentially negatively and positively biased patient test results for HbA1c. The firm is investigating the root cause.

Source facts

Action
On May 22, 2019 the firm issued the Product Recall letter along with a customer reply form to all customers informing them that the firm has identified that the calibrator level 2 may have been manufactured incorrectly for a portion of the lots and that although this lot may generate an active calibration curve, a shift to QC and patient results may occur. Between concentration range of 5.33% to 6.87% there is no impact to assay performance, however, outside of this range, a bias of greater than 3% may be seen with patient results. For US customers the Product Recall Letters were sent via FedEx Priority Overnight Express (note-for OUS, each country organization will establish their own method of notification, which consistence of mailings or customer visits). The Product Recall Letter instructed customers to: - Immediately discontinue use of the calibrator lot number and switch to the alternate calibrator lot. - Destroy any remaining inventory of impacted material according to customer's laboratory procedures. - Review this letter with Medical Director or Laboratory Management and follow laboratory protocol regarding the need for reviewing previously reported patient results. - Complete and return the Customer Reply Form. Contact your local firm representative for assistance with replacement product. - If customers have forwarded the product to other laboratories, please inform them of this Product Recall and provide them with a copy of this letter. - Please retain this letter for your laboratory records. If any health care providers have any questions regarding this information, U.S. Customers please contact Customer Service at 1-877-4ABBOTT (available 24 hours a day, 7 days a week). Customers outside the U.S., please contact your local area Customer Service. If any patient or user injury are experienced associated with this Field Action, please immediately report the event to your local area Customer Service.
Initiated
2019-05-22
Root cause
Under Investigation by firm
Terminated
2021-01-22
Distribution
Worldwide distribution - US Nationwide distribution in the states of AR, AZ, CA, FKL GA, IA, ID, IL, IN, KS, LA, MN, MO, NC, NE, NJ, NV, NY, OH, OK, PA, SC, SD, TX, UT, VA, and WI; and countries of Austria, Germany, Italy, South Africa, and Spain.
Recall number
Z-1688-2020
Recall status
Terminated
CDRH record ID
180359
FDA record created
2020-04-09
Recall event number
85247

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗