Open original FDA record ↗Recall record
- FDA identifier
- Z-1688-2021
- Company in FDA record
- Ortho-Clinical Diagnostics, INc.
- FDA record creation date
- May 27, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Ortho investigated and confirmed the potential for unexpected failed calibration, and condition codes associated with the calibration, when using VITROS SARS-CoV-2 Ag Reagent Pack along with VITROS Immunodiagnostic Products SARS-CoV-2 Ag Calibrator, Lot 20.
Source facts
- Action
- On May 4, 2021, Ortho Clinical Diagnostics issued an Urgent Medical Device Correction notice to customers via letter notifying them of calibration failures and the associated condition code for specified calibrator lots. Customers received the following instructions. " Discontinue using VITROS SARS-CoV-2 Antigen Reagent/Calibrator Lot 0064, discard any remaining inventory and switch to an alternate lot. " Ortho will credit and/or replace all remaining inventory of VITROS SARS-CoV-2 Antigen Reagent/Calibrator Lot 0064 as indicated on the Confirmation of Receipt form. " Complete the enclosed Confirmation of Receipt form no later than MM/DD/YY, even if your facility no longer has the affected product in inventory. " Please forward this notification if the product was distributed outside of your facility.
- Initiated
- 2021-05-03
- Root cause
- Under Investigation by firm
- Terminated
- 2023-10-24
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of AR, CA, IA, IL, LA, MO, MS, TX, WI, and the countries of Colombia, India, Japan, Italy, Spain, Portugal, France.
- Recall number
- Z-1688-2021
- Recall status
- Terminated
- CDRH record ID
- 187438
- FDA record created
- 2021-05-27
- Recall event number
- 87888