Skip to content

Public FDA record

Z-1688-2024: Brand Name: PIEZON Bottle Product Name: PIEZON Bottle Model/Catalog Number: EG-111 Software Version: N/A

Recall record Z-1688-2024 for Electro Medical Systems SA. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1688-2024
Company in FDA record
Electro Medical Systems SA
FDA classification date
Apr 24, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Received complaints that bottles assembled with one batch of bottle collars can be disassembled in two parts when used under pressure (during treatment or storage if the use is not releasing the pressure inside the bottles).

Source facts

Action
On 3/13/2024, the firm notified, via email, its distributors and asked them to extend the recall to end users. Currently the Customer Notification Letter to end users is being drafted (date of distribution pending). The Customer Notification Letter will information customers that the firm is recalling WATER, CLEANER and PIEZON Bottles due to a manufacturing issue that may result with these affected bottles to leak and/or break and in unfavorable cases may lead to injuries. Customers are instructed to: 1. Cease use of the affected products. 2. Return the affected products to their distributors For questions or assistance, contact EMS Dallas at 1-800-367-0367 or email to emsrepairs@ems-na.com Monday thru Friday 8:00 a.m. to 5:00 p.m. CST.
Initiated
2024-03-13
Root cause
Process control
Distribution
Worldwide distribution - U.S. Nationwide in the states of IL and TX. The countries of ALBANIA, ARMENIA, AUSTRALIA, AZERBAIJAN, BAHRAIN, BELGIUM, BRAZIL, BULGARIA, CANADA, CHINA, CZECH REPUBLIC, EGYPT, FRANCE, GEORGIA, GERMANY, GREECE, HONG KONG, HUNGARY, INDONESIA, ISRAEL, ITALY, JAPAN, JORDAN, KAZAKHSTAN, KENYA, KOREA REPUBLIC OF, KOSOVO, KUWAIT, LUXEMBOURG, MALAYSIA, MEXICO, MOLDOVA REPUBLIC OF, NETHERLANDS, NEW ZEALAND, OMAN, PALESTINE, PHILIPPINES, POLAND, QATAR, ROMANIA, RUSSIAN FEDERATION, SAUDI ARABIA, SINGAPORE, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TURKEY, Thailand, UKRAINE,UNITED ARAB EMIRATES, UNITED KINGDOM, UZBEKISTAN, VIETNAM.
Recall number
Z-1688-2024
Recall status
Open, Classified
CDRH record ID
206855
FDA classified
2024-04-24
Recall event number
94293

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.