Z-1690-2024: Philips Patient Information Center (PIC) iX Uninterruptable Power Supply (UPS): 1. APC (Schneider Electri

Recall record Z-1690-2024 for Philips North America Llc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1690-2024
Company in FDA record
Philips North America Llc
FDA classification date
Apr 25, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Possible failure of Uninterruptable Power Supply (UPS) devices supporting the Patient Information Center (PIC) iX system, potential for the associated PIC iX system to shut down due to the lack of power. may lead to the delay in the detection of a change in a patient condition

Source facts

Action
Philips issued Urgent Medical Device Correction Letter on 3/22/24. Letter states reason for recall, health risk and action to take: If your UPS fails : 1) Promptly take the UPS offline and plug any devices that had been attached to it into a different appropriate power source. Note: It is recommended that UPS devices intended for operation in network and server rooms be kept in a temperature and humidity controlled environment, ensuring adequate airflow around the UPS to prevent UPS device failure. This notice should be passed on to all those who need to be aware within your organization or to any organization where affected devices have been transferred. 5. Actions planned by Philips to correct the problem A Philips representative will contact you to schedule a visit from a Philips Field Service Engineer who will replace the malfunctioning UPS. If you need any further information, please contact your local Philips representative, or call 1-800- 722- 9377. Philips Reference# 2024-CC-HPM-010
Initiated
2024-03-22
Root cause
Component design/selection
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Aruba, Australia, Austria, Belgium, Bolivia, Canada, Chile, China, Colombia, Costa Rica, Czech Republic, Ecuador, Estonia, Fiji, Finland, France, Germany, Greece, Hong Kong, India, Indonesia, Iran, Ireland, Israel, Italy, Kazakhstan, Kenya, Kuwait, Latvia, Lebanon, Malaysia, Mexico, Myanmar, Netherlands, New Zealand, Oman, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Saudi Arabia, Singapore, South Africa, Spain, Switzerland, Taiwan, United Kingdom, Utd. Ara Emir.
Recall number
Z-1690-2024
Recall status
Open, Classified
CDRH record ID
206965
FDA classified
2024-04-25
Recall event number
94340

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗