Open original FDA record ↗Recall record
- FDA identifier
- Z-1690-2026
- Company in FDA record
- Medline Industries, LP
- FDA classification date
- Mar 27, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Connectors were not manufactured to required dimensional specifications and may not form a proper seal with ENFit-style devices, which could lead to leakage. Leakage may result in delayed care and fluid loss. Additional consequences may also occur if user over-tightens the connector to stop the leakage, including additional clinical intervention, risk of infection, patient discomfort, and prolonged recovery process.
Source facts
- Action
- On February 13, 2026, the firm notified affected consignees via Urgent Medical Device Recall letters delivered through First Class mail and email. Customers were instructed to destroy all affected ENFit G-Tube connectors and kits that contain the G-Tube connectors.
- Initiated
- 2026-02-13
- Root cause
- Process change control
- Distribution
- Worldwide - US Nationwide and the countries of Canada.
- Recall number
- Z-1690-2026
- Recall status
- Open, Classified
- CDRH record ID
- 218983
- FDA classified
- 2026-03-27
- Recall event number
- 98524