Skip to content

Public FDA record

Z-1690-2026: ENFIT G-TUBE CONNECTOR, MEDLINE Item No. ENFIT1010GC

Recall record Z-1690-2026 for Medline Industries, LP. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1690-2026
Company in FDA record
Medline Industries, LP
FDA classification date
Mar 27, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Connectors were not manufactured to required dimensional specifications and may not form a proper seal with ENFit-style devices, which could lead to leakage. Leakage may result in delayed care and fluid loss. Additional consequences may also occur if user over-tightens the connector to stop the leakage, including additional clinical intervention, risk of infection, patient discomfort, and prolonged recovery process.

Source facts

Action
On February 13, 2026, the firm notified affected consignees via Urgent Medical Device Recall letters delivered through First Class mail and email. Customers were instructed to destroy all affected ENFit G-Tube connectors and kits that contain the G-Tube connectors.
Initiated
2026-02-13
Root cause
Process change control
Distribution
Worldwide - US Nationwide and the countries of Canada.
Recall number
Z-1690-2026
Recall status
Open, Classified
CDRH record ID
218983
FDA classified
2026-03-27
Recall event number
98524

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗