Reason stated in the source
GE Healthcare has become aware of an Air Kerma Rate issue associated with a large Equivalent Patient Thickness (EPT) on the Innova system that my impact patient safety. During a Maximum Air Kerma Rate (AKR) calibration procedure check, GE Healthcare found the AKR calibration procedure does not exactly reflect the worst-case clinical usage of the product. This may result, when using large Equival
Source facts
- Action
- Consignees were sent a GE Healthcare "Important electronic Product Radiation Warning" letter dated November 3, 2008. The letter was addressed to Healthcare Administrator/Risk Managers, Managers of Radiology/Cardiology, Radiologists/Cardiologists. The letter described the Defect Description and Related Hazards, Affected Product Details, Safety Instructions, Product Correction and Contact Information. For further information, contact the GE Healthcare Technical Support Center at 1-800-437-1171.
- Initiated
- 2008-10-09
- Root cause
- Radiation Control for Health and Safety Act
- Terminated
- 2012-01-07
- Distribution
- Nationwide Distribution -- United States including Puerto Rico and Guam.
- Recall number
- Z-1691-2009
- Recall status
- Terminated
- CDRH record ID
- 85121
- FDA classified
- 2009-09-24
- Recall event number
- 53260