Open original FDA record ↗Recall record
- FDA identifier
- Z-1691-2016
- Company in FDA record
- Sedecal USA, Inc.
- FDA record creation date
- Jun 3, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Due to a software defect, the system may sporadically apply the default x ray exposure parameters for an adult ( patient type : Normal ) even though the patient type " Newborn" was selected and is displayed in the generator control area of the Eleva User Interface.
Source facts
- Action
- Sedecal USA has developed a new software release to address this issue. The distributor of this product, Philip, will modify the systems in the field to correct the defect at no cost to the customer. 1. You will contact customers and initiate a software update to correct the two defects. 2. You will install the software and perform testing to ensure the software update was effective. 3. The customer notification letter which includes a statement that you will without charge, remedy the defect or bring the product into compliance. For further questions, please call (847) 394-6960.
- Initiated
- 2015-03-23
- Root cause
- Radiation Control for Health and Safety Act
- Terminated
- 2018-03-09
- Distribution
- US Distribution - TX, OH, NY, MD, IN, PA, DE, WA, NH, AZ, GA, NC, UT, VT, CA, LA, NJ, MN, FL, WI, HI, IA, MA, ME, TN, CT, ID, MI, AR and KS.
- Recall number
- Z-1691-2016
- Recall status
- Terminated
- CDRH record ID
- 146293
- FDA record created
- 2016-06-03
- Recall event number
- 74106