Z-1691-2017: Altius M-INI OCT Posterior Spinal Fixation System, Model Number 1200-9216 The AltiusTM OCT and AltiusTM M

Recall record Z-1691-2017 for Zimmer Biomet Spine, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1691-2017
Company in FDA record
Zimmer Biomet Spine, Inc
FDA record creation date
Mar 28, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Zimmer Biomet is recalling the Altius M-INI OCT Posterior Spinal Fixation System because the driver is unable to mate with the screw head tulip.

Source facts

Action
Zimmer Biomet sent an Urgent Medical Device Recall letter on March 20, 2017, to all affected consignees. The letter identified the product, the problem and the action to be taken by the consignee. Consignee's Responsibilities 1. Review this notification and ensure affected team members are aware of the contents. 2. Prior to sending Altius Kits to central processing, perform a functional check of the driver withall screws in the kit to ensure they mate. If the drivers do not mate, please return in accordance with Step 3b. 3. Complete the Certification of Acknowledgement portion of Attachment 1 Inventory Return Certification Form Phase I. b. If the driver is unable to mate with the screws in the set, immediately sterilize and return the affected nonconforming product from your distributorship and affected hospitals within your territory along with a completed Attachment 1 Inventory Return Certification Form Phase I and Attachment 2 Certificate of Sterilization to Zimmer Biomet. i. For each return, send a copy of Attachment 1 to jeremy.hansmann@zimmerbiomet.com ii. Include a hardcopy of Attachment 1 with your shipment for immediate processing. iii. Include a copy of Attachment 2 Certificate of Sterilization with returned instruments. iv. Mark the outside of the returns box(es) clearly with RECALL. 4. Distributors will notify their accounts to which the product was further distributed and to locate and return products from these accounts. Using the Additional Accounts Form provided with the email notice sent to your facility, return contact information for any additional hospitals that have the affected product. If there are no additional users to notify, please confirm this by annotating and returning the form provided indicating that no additional users have been identified. 5. Retain a copy of your field action acknowledgement and product return forms for your records in the event of a compliance audit of your facility. 6. If
Initiated
2017-03-16
Root cause
Device Design
Terminated
2018-07-30
Distribution
Worldwide Distribution - US (nationwide) and Internationally to the Netherlands
Recall number
Z-1691-2017
Recall status
Terminated
CDRH record ID
153660
FDA record created
2017-03-28
Recall event number
76610

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗