Connectors were not manufactured to required dimensional specifications and may not form a proper seal with ENFit-style devices, which could lead to leakage. Leakage may result in delayed care and fluid loss. Additional consequences may also occur if user over-tightens the connector to stop the leakage, including additional clinical intervention, risk of infection, patient discomfort, and prolonged recovery process.
Source facts
Action
On February 13, 2026, the firm notified affected consignees via Urgent Medical Device Recall letters delivered through First Class mail and email. Customers were instructed to destroy all affected ENFit G-Tube connectors and kits that contain the G-Tube connectors.
Initiated
2026-02-13
Root cause
Process change control
Distribution
Worldwide - US Nationwide and the countries of Canada.
Recall number
Z-1691-2026
Recall status
Open, Classified
CDRH record ID
218984
FDA classified
2026-03-27
Recall event number
98524