Wiring in battery back-up power supply may short and cause thermal damage to ventilator.
Source facts
Action
All customers were notified by Federal Express letter and were contacted by phone through field service engineers. They were told to examine their inventory for any installed or spare part battery assemblies and to remove and quarantine these parts. Any ventilators with installed affected battery assemblies were to be taken out of service.
Initiated
2008-03-21
Root cause
Process control
Terminated
2008-10-29
Distribution
Nationwide, including one VA facility in IL and two military facilities in NC and NY. No foreign distribution.
Recall number
Z-1692-2008
Recall status
Terminated
CDRH record ID
69664
FDA classified
2008-09-16
Recall event number
47523