PerkinElmer Life and Analytical Sciences, Wallac, OY
FDA record creation date
May 4, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II
Reason stated in the source
There is a potential for leaking vials for certain lot numbers where, if proper attire is not used as instructed, a transient minor allergic skin reaction may result when handling leaking succinylacetone reagent vials.
Source facts
Action
Actions to be taken:
- Inspect your inventory of the NeoBase Succinylacetone Assay Solution vials for the affected lots.
- If you notice leakage or deformed stopper seals, destroy the affected vials according to local instructions and regulations.
- Complete the Response Form with the quantity and identity of affected Succinylacetone Assay Solution vials in your
inventory.
- Return the Response Form to PerkinElmer.
- PerkinElmer will replace the quantity of leaking Succinylacetone Assay Solution vials found and destroyed from your inventory
Initiated
2017-11-02
Root cause
Process control
Terminated
2019-09-17
Distribution
Distributed in 20 US states: AL, AR, AZ, CA, CO, FL, GA, IL, IN, KS, KY, LA, MI, MO, NV, OH, OR, SC, TN, TX.
Recall number
Z-1692-2018
Recall status
Terminated
CDRH record ID
161848
FDA record created
2018-05-04
Recall event number
79303