Z-1692-2018: NeoBase Succinylacetone Assay Solution, Product Number: 3042-0020; Affected UDI Numbers: (01)064381472957

Recall record Z-1692-2018 for PerkinElmer Life and Analytical Sciences, Wallac, OY. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1692-2018
Company in FDA record
PerkinElmer Life and Analytical Sciences, Wallac, OY
FDA record creation date
May 4, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a potential for leaking vials for certain lot numbers where, if proper attire is not used as instructed, a transient minor allergic skin reaction may result when handling leaking succinylacetone reagent vials.

Source facts

Action
Actions to be taken: - Inspect your inventory of the NeoBase Succinylacetone Assay Solution vials for the affected lots. - If you notice leakage or deformed stopper seals, destroy the affected vials according to local instructions and regulations. - Complete the Response Form with the quantity and identity of affected Succinylacetone Assay Solution vials in your inventory. - Return the Response Form to PerkinElmer. - PerkinElmer will replace the quantity of leaking Succinylacetone Assay Solution vials found and destroyed from your inventory
Initiated
2017-11-02
Root cause
Process control
Terminated
2019-09-17
Distribution
Distributed in 20 US states: AL, AR, AZ, CA, CO, FL, GA, IL, IN, KS, KY, LA, MI, MO, NV, OH, OR, SC, TN, TX.
Recall number
Z-1692-2018
Recall status
Terminated
CDRH record ID
161848
FDA record created
2018-05-04
Recall event number
79303

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗