Z-1692-2026: Vue Motion V12. Product Number: 1017979.

Recall record Z-1692-2026 for PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1692-2026
Company in FDA record
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
FDA classification date
Mar 27, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential that mis-ordered frames in Vue Motion during dynamic cine runs may cause images frames to display out of sequence.

Source facts

Action
Philips Medical notified consignees on about 03/02/2026 via URGENT Medical Device Correction letter. Consignees were instructed to first identify, and how to identify, if the affected products are on hand. If so, they were further instructed that if they do NOT have thumbnail enabled, they will NOT experience this issue described, but if they HAVE thumbnail enabled, when reviewing multi frame studies, always review the frame number displayed on the screen as the correct frame number is visible on the screen. Consignees were told that they may continue to use your system(s) in accordance with the intended use and by provided the recommendation in the letter, to circulate the letter to all users of this device so that they are aware of the potential issue retain this letter with your system(s) until a solution is installed on your system; ensure the letter is in a place likely to be seen/viewed. Consignees were also requested to complete and return the provided response form. A Philips representative will contact each consignee to schedule a time to install a software solution on the affected system(s) to resolve the issue (reference C&R 2025-EI-RI-003).
Initiated
2026-03-05
Root cause
Software design
Distribution
Worldwide distribution - US Nationwide and the countries of Algeria, Andorra, Angola, Argentina, Australia, Austria, Belgium, Bolivia, Brazil, Cambodia, Canada, Chile, China, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Finland, France, Germany, Gibraltar, Greece, Guernsey, Holy See (Vatican City State), Honduras, Hungary, Iceland, India, Indonesia, Ireland, Israel, Italy, Japan, Jersey, Jordan, Kuwait, Lebanon, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Norway, Oman, Palestine, State of, Panama, Papua New Guinea, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, United Arab Emirates, United Kingdom, Uruguay, Viet Nam, Zambia.
Recall number
Z-1692-2026
Recall status
Open, Classified
CDRH record ID
218977
FDA classified
2026-03-27
Recall event number
98520

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗