Z-1693-2015: Neusoft NeuViz 64 Multi-slice CT Scanner System

Recall record Z-1693-2015 for Neusoft Medical System Co. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1693-2015
Company in FDA record
Neusoft Medical System Co
FDA classification date
Jun 16, 2015
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The A-plane compensator is made of Teflon which will become fragile due to X-ray exposure passing through it. The thinnest part of the compensator is subject to radiation degradation and as a result has been found to fail prematurely. Failure is indicated by pin or ring artifact in the image

Source facts

Action
Neusoft planned action: 1. Notification of all dealers and purchasers is to be made within 15 working days of receipt of this letter in the manner specified in 21 CFR 1003.21 and 1003.22. This office and the Food and Drug Administration (FDA) district office coordinator noted below are to be included in the notification. 2. Neusoft Medical Systems Co., Ltd will implement actions to monitor the exposures of the tube and when the exposure number is close to 300,000 seconds, the A-plane compensator will be replaced. The Food and Drug Administration (FDA) district office coordinator should be notified when the last A-plane compensator is replaced. 3. The Customer Information Letter has been sent to customers as a corrective action. The maximum exposure number of 5 US systems is 191830 seconds as of Apr.20, 2015, so the firm will monitor the data continuously and replace the A-plane compensators before the defined threshold. 4. Neusoft will replace all compensators that have exceeded the 300,000 seconds free of charge and will replace all compensators that fail during warranty and the compensators failed less than 300,000 seconds though out of warranty free of charge. You may proceed or continue with implementation of the CAP (subject to the conditions noted above) For further questions please call (281) 453-1205
Initiated
2014-03-08
Root cause
Radiation Control for Health and Safety Act
Terminated
2016-12-05
Distribution
Worldwide Distribution - US Distribution including Puerto Rico and to the states of : IL, MA, TX and SC., and to the countries of : China, Malaysia, Lebanon, Ecuador, Egypt, United Arab Emirates,Kazashstan, Tunisia, Angola, Brazil, Uzbekistan, Lebanon, Cote D' ivoire, South Africa, Montenegro and Tajikistan.
Recall number
Z-1693-2015
Recall status
Terminated
CDRH record ID
137271
FDA classified
2015-06-16
Recall event number
71306

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗