Open original FDA record ↗Recall record
- FDA identifier
- Z-1693-2017
- Company in FDA record
- Baxter Healthcare Corporation
- FDA record creation date
- Oct 2, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
SIGMA Spectrum Infusion Pumps with Master Drug Library, Software V8.00.02 contains an anomaly that may cause a Sharp Watchdog Timeout error message. Specifically, the software anomaly causes the error message only under conditions in which a message is being written to the Spectrum Infusion Pumps Event History Log (EHL), and the length of that message exactly matches the size of the available memory at the end of a memory sector. For this reason, the issue is significantly more likely to manifest when poor network maintenance results in high volumes of messages being written to the EHL.
Source facts
- Action
- URGENT MEDICAL DEVICE CORRECTION Letters and Reply Forms (dated 1/26/2017) were mailed to customers via USPS first-class mail on 1/26/2017 providing the necessary instructions to customers. For questions, please contact your Baxter sales representative, or Baxter Healthcare Medina at 800-356-3454 (choose option 1) Monday through Friday, between 7:00 am and 7:00 pm ET.
- Initiated
- 2017-01-26
- Root cause
- Process design
- Terminated
- 2018-05-24
- Distribution
- Natiowide, Puerto Rico and Canada
- Recall number
- Z-1693-2017
- Recall status
- Terminated
- CDRH record ID
- 152976
- FDA record created
- 2017-10-02
- Recall event number
- 76364