Open original FDA record ↗Recall record
- FDA identifier
- Z-1693-2020
- Company in FDA record
- Medtronic Navigation, Inc.
- FDA classification date
- Apr 10, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Product was incorrectly assembled which could affect navigation accuracy of the device.
Source facts
- Action
- On 08/28/2019, the firm sent an "URGENT: MEDICAL DEVICE RECALL" Notification to its customer informing them of the manufacturing issue (incorrect assembly) and instructed them to: 1. Examine their inventory, and immediately quarantine the affected NavLock" Tracker and to return to the Recalling Firm. 2. To contact the Recalling Firm at 1-888-826-5603 to receive a return material authorization (RMA) and to arrange for return and a no-charge replacement. Once an RMA number is obtained, ship the affected product to Medtronic Product Services, 1480 Arthur Ave, Louisville CO 80027. 3. Follow instructions on the attached Customer Confirmation Form. Sign and date the bottom of the form and then return the form to RS.NavFCA@medtronic.com or fax it to Medtronic at 651-367-7075.
- Initiated
- 2019-08-28
- Root cause
- Employee error
- Terminated
- 2021-05-05
- Distribution
- U.S.: NY No O.U.S.
- Recall number
- Z-1693-2020
- Recall status
- Terminated
- CDRH record ID
- 180507
- FDA classified
- 2020-04-10
- Recall event number
- 85318