Open original FDA record ↗Recall record
- FDA identifier
- Z-1694-2009
- Company in FDA record
- X Spine Systems Inc
- FDA classification date
- Aug 19, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
A defect cause was discovered following a customer complaint pertaining to a post operative dissociation of the Capless Li Screw construct. The screw assembly is made up of three parts; a cup which locks onto a 5.5mm rod, a yoke that cradles the rod during fusion, and a screw that attached to the vertebra. The screw ball on the screw was manufactured out of tolerance (undersized).
Source facts
- Action
- X Spine Systems Inc. issued advisory notices in March 2009 to Consignees notifying them of the problem and requesting all affected product be returned to the firm. The firm also conducted telephone calls and emails for the return of the product. For further information, contact X Spine Systems Inc. at 1-937-847-8400 extension 115.
- Initiated
- 2009-03-24
- Root cause
- Process control
- Terminated
- 2012-04-25
- Distribution
- Worldwide Distribution -- Including states of AZ, CA, CO, FL, ID, IN, MO, MS, NV, OK, TN, TX, UT and WY and countries of Spain, South Africa and the United Kingdom.
- Recall number
- Z-1694-2009
- Recall status
- Terminated
- CDRH record ID
- 81734
- FDA classified
- 2009-08-19
- Recall event number
- 51956