Open original FDA record ↗Recall record
- FDA identifier
- Z-1694-2013
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA classification date
- Jul 11, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Firm became aware of an unintended behavior when using the Mammomat Inspiration. The possibility exists that a malfunction may occur during Stereo or Tomo acquisitions if the tomo clutch was incorrectly adjusted. Although the possibility is rare, the system could move while a patient's breast is compressed resulting in a possible bruise. Firm released Update Instructions XP031/12/S for this iss
Source facts
- Action
- Siemens issued an Update Instructions XP031/12/S by letter to the affected customers. For affected sites, a Siemens Service Engineer will inspect the unit and make any adjustments needed. Following the inspection and adjustments, a proper calibration will be completed. For questions call 610-448-3237 and 610-446-4634.
- Initiated
- 2013-02-28
- Root cause
- Device Design
- Terminated
- 2014-02-21
- Distribution
- United States Nationwide Distribution in the states of ND, OH and SC.
- Recall number
- Z-1694-2013
- Recall status
- Terminated
- CDRH record ID
- 119086
- FDA classified
- 2013-07-11
- Recall event number
- 65487