Skip to content

Public FDA record

Z-1694-2026: Brand Name: Olympus SOLTIVE Premium SuperPulsed Laser System Product Name: SOLTIVE Premium Laser System T

Recall record Z-1694-2026 for Olympus Corporation of the Americas. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1694-2026
Company in FDA record
Olympus Corporation of the Americas
FDA classification date
Mar 30, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Following a complaint investigation, Olympus identified a potential defect in the 24V power supply module in some SOLTIVE laser units which may cause the system to become inoperable. Additionally, smoke or a burning smell may occur. By design, the issue causing the smoke or burning smell would be contained within the internal laser console enclosure and would be self-extinguishing.

Source facts

Action
On February 27, 2026 URGENT: MEDICAL DEVICE CORRECTION letters were sent to customers. Actions to be taken: 1. Examine your inventory and identify any of the affected devices listed in the attachment. 2. Olympus requests that you acknowledge receipt of this letter through our recall web portal. 3. Olympus representatives will begin contacting customers to coordinate the service of affected unit(s). In the meantime, you may continue using your SOLTIVE Laser until the service is completed. Olympus will replace the 24V power supply module at no charge to you. Note: Any issues beyond the replacement of the 24V module in this corrective action will follow the standard SOLTIVE Laser repair process and your service contract, if applicable. 4. If you have further distributed the affected products, please forward this notice to other users who may be impacted.
Initiated
2026-02-27
Root cause
Process control
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Chile, Europe, Singapore, Australia, and Hong Kong.
Recall number
Z-1694-2026
Recall status
Open, Classified
CDRH record ID
218671
FDA classified
2026-03-30
Recall event number
98422

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗