A calibrated torque wrench malfunctioned at the supplier leading to possible incorrect torque of x-ray tube mounting fasteners.
Source facts
Action
Philips sent an Urgent - Field Safety Notice dated April 13, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer.
Philips Healthcare is recommending to stop the use of your system until the X-ray tube mounting fasteners are replaced.
In addition, a Philips Field Service Engineer (FSE) will service your system via a
Mandatory Field Change Order (FCO 72800638). A Philips FSE will contact you
to schedule this appointment.
If you need any further information or support concerning this issue, please contact your local Philips representative or your local Philips Healthcare office.
This notice has been submitted to the appropriate Regulatory Agency.
Philips apologizes for any inconveniences this problem has caused your organization. For further questions please call (440) 483-7600.
Initiated
2015-04-13
Root cause
Other
Terminated
2015-10-01
Distribution
Internationally to the countries of China and Norway.
Recall number
Z-1695-2015
Recall status
Terminated
CDRH record ID
136899
FDA classified
2015-05-29
Recall event number
71231