- Recall classification
- Class II
Reason stated in the source
The detector may signal that it is ready for acquisition when it actually is not, resulting in failure to properly acquire the X-ray image.
Source facts
- Action
- Philips Healthcare plans to correct the defect at no cost to customers by implementing a field correction involving the installation of new software. 1. You will contact customers and initiate a software update to correct the defect. 2. You will install the software and perform testing to ensure the software update was effective. 3. The customer notification letter which includes a statement that you will without charge, remedy the defect or bring the product into compliance. For further questions, please call 1-800- 722-9377.
- Initiated
- 2015-11-10
- Root cause
- Radiation Control for Health and Safety Act
- Terminated
- 2017-05-03
- Distribution
- USA (nationwide) Distribution to the states of : ( DuraDiagnost) to the states of : OH, CA, WA and NE. (DigitalDiagnost) to the states of : AK, AZ, AL, CA, CT, CO, DE, FL, GA, HI, ID, KY, LA, MA, MN, MD, MS, ME, MO, NE, NY, ND, OH and PA.
- Recall number
- Z-1695-2016
- Recall status
- Terminated
- CDRH record ID
- 146236
- FDA classified
- 2016-06-03
- Recall event number
- 74083