Skip to content

Public FDA record

Z-1695-2023: RR Mechatronics Control Level A ESR Control4.6 ml QRR049002- In-vitro diagnostic quality control materi

Recall record Z-1695-2023 for Mechatronics USA. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1695-2023
Company in FDA record
Mechatronics USA
FDA classification date
Jun 7, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class III

Reason stated in the source

QRR049001 Lot QCAB6AN505 incorrectly labelled as QRR049002 Lot QCAB6AAA28. The incorrectly labelled vial will upload the range for the Starrsed Control Level A and the instrument result will be flagged as out of range.

Source facts

Action
Mechatronics issued Urgent Field Safety corrective action letter on 4/14/23 via email. Letter states reason for recall, health risk and action to take: Field Safety Field corrective action to be taken by the customer: 1. Confirm that you have received this information by returning enclosed acknowledgement form. 2. Please check your inventory for QRR049001-Starrsed Control Level N-Lot QCAB6AN505. 3. Please check your inventory for QRR049002-Starrsed Control Level A-Lot QCAB6AAA28. 4. Confirm the date and number of unused incorrectly labeled vials in your inventory. 5. Discard any unused incorrectly labelled vials in your inventory. 6. Please indicate the number of requested replacement vials. Please document the information on the enclosed Return Form and send to sales@rrmechatronics.com Transmission of this notice We kindly ask you to inform those who need to be aware of this notification within your organization or any other organization to which the affected product(s) have been transferred. We need confirmation of the complete delivery of this lot. By completing the above-mentioned enclosed Return Form by yourself and affected customers.
Initiated
2023-04-14
Root cause
Labeling Change Control
Distribution
Worldwide distribution - US Nationwide distribution in the state of OH and the country of Netherlands.
Recall number
Z-1695-2023
Recall status
Open, Classified
CDRH record ID
200049
FDA classified
2023-06-07
Recall event number
92276

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗