Skip to content

Public FDA record

Z-1695-2024: Abbott Aveir Leadless Pacemaker, REF: LSP112V, STERILEEO

Recall record Z-1695-2024 for Abbott Medical. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1695-2024
Company in FDA record
Abbott Medical
FDA classification date
Apr 29, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Their is a potential that electromagnetic interference may cause leadless pacemaker may result in mode change.

Source facts

Action
On April 3, 2024, Abbott issued a "Urgent Medical Device Correction" to affected consignees via certified mail, email or hand-delivery. Abbott asked consignees to take the following actions: 1. Prophylactic device replacement is NOT recommended. o All currently manufactured LSP112V devices utilize the upgraded firmware. o Following the firmware upgrade, the implanted device will be equivalent to newly manufactured LSP112V devices. 2. As part of follow-up, suggested within 3 months, upgrade the LSP112V firmware. o For most devices, the upgrade will execute automatically when interrogated. If required, contact Technical Services to assist with the upgrade. o If the device presents in MRI or EVVI mode, reprogram the device to the desired mode and settings. 3. Please return a completed Acknowledgement Form and maintain a record of this notice . along with a copy of the completed Acknowledgement Form to ensure effectiveness of the communication. 4. Adverse reactions or quality problems experienced with the use of this product should be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail or by fax. - Complete and submit the report Online: www.fda.gov/medwatch/report.htm - Regular Mail or Fax: Download form https://www.fda.gov/safety/medical-product-safety-information/medwatchforms-fda-safety-reporting or call 1-800-332-1088 to request a reporting form, then complete and return to the address on the pre-addressed form, or submit by fax to 1-800-FDA-0178
Initiated
2024-02-05
Root cause
Device Design
Distribution
US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV. OUS: Pending
Recall number
Z-1695-2024
Recall status
Open, Classified
CDRH record ID
206245
FDA classified
2024-04-29
Recall event number
94072

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗