Z-1696-2021: Cascadia AN Interbody Convex 10x22x15mm Catalog Number: 6101-2102215NC-G2 Lumbar implants are interver

Recall record Z-1696-2021 for K2M, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1696-2021
Company in FDA record
K2M, Inc
FDA record creation date
Jun 1, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Mislabeled product labeling references incorrect length or height, or length and height for specific Lots, Sizes 10x22x14mm, 10x28x14mm, 10x22x15mm

Source facts

Action
Stryker Spine issued an Urgent Medical Device Recall letter to Stryker Spine Branches/Agencies/Distributors/Sales Representatives via e-mail on May 7, 2021 and instructed to identify and return all 3 lots of product to Stryker Spine. Hospital notified by overnight mail (FedX) on 5/10/21. Letter states reason for recall and action to take: 1.Our records indicate product associated with the above catalog and lot numbers were previously distributed to your hospital. Please note this product has already been removed by your local Stryker Sales Representative and/or Stryker Branch or Agency. This communication is being made for notification purposes only; there are no additional product(s) to return for this matter. Stryker has contained all product associated with this recall. 2.Please assist us in meeting our regulatory obligation by completing and signing the enclosed Business Reply Form and email a copy to Stryker Spine at SpineRegulatoryActions@Stryker.com, or fax the form to 201.575.4675. 3.Maintain awareness of this communication internally within your facility. If you have any questions or concerns, please contact Stryker Regulatory Compliance at 201.749.8090. OUS: Europe is the only region outside of the US that received parts from these lots. All parts in Europe have been quarantined at Stryker Europe locations (Netherlands and Italy) prior to distribution to customers. Accordingly, there are no notifications required for Europe
Initiated
2021-05-07
Root cause
Labeling Change Control
Terminated
2023-03-08
Distribution
Worldwide distribution - US Nationwide distribution in the states of MA, MI, NJ, PA, UT and the countries of Italy, Netherlands.
Recall number
Z-1696-2021
Recall status
Terminated
CDRH record ID
187395
FDA record created
2021-06-01
Recall event number
87905

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗