Skip to content

Public FDA record

Z-1697-2025: Brand Name: AirLife Product Name: AirLife Infant Heated Wire Circuit Model/Catalog Number: AH165 Software

Recall record Z-1697-2025 for Vyaire Medical. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1697-2025
Company in FDA record
Vyaire Medical
FDA classification date
May 9, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Adapters may disconnect during setup or while in use, potentially interrupting ventilation. Any disconnection can result in severe consequences for neonates, including hypoxia, hypercapnia, and organ failure.

Source facts

Action
On April 10, 2025, the firm notified customers via emailed letters titled URGENT: MEDICAL DEVICE RECALL. Customers are instructed to examine their inventory for the affected lots of product. Product in inventory should be quarantined. For in-use product - 1) For affected product in use that is utilizing the adaptor connection, immediately stop use of the product. 2) For affected product in use that is not utilizing the adaptor connection, ventilator circuit may continue to be utilized. Customers were provided with additional precautions to take to ensure proper function. Customers can choose between returning or destroying affected product. Firm will provide replacement product. If you have any questions regarding this field action, please call AirLife at 1-800-433-2797, or e-mail at productquality@myairlife.com.
Initiated
2025-04-10
Root cause
Component design/selection
Distribution
Worldwide distribution - US Nationwide and the countries of Germany, Poland, Belgium, Spain, Romania, UK, Italy, Netherlands, India, Slovenia, Qatar, and Saudi Arabia.
Recall number
Z-1697-2025
Recall status
Open, Classified
CDRH record ID
213370
FDA classified
2025-05-09
Recall event number
96664

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗