Skip to content

Public FDA record

Z-1698-2024: ADVANCED RESEARCH MEDICAL Lumbar Interbody Fusion System (OLLIF), Flexible Curette Loop Blade, Part # 01-

Recall record Z-1698-2024 for Advanced Research Medical, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1698-2024
Company in FDA record
Advanced Research Medical, LLC
FDA classification date
Apr 29, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential shearing of the disposable blade may occur and be contained within the of Flexible Curette Shaft Assembly during surgery.

Source facts

Action
The sole distributor was notified by emailed letter of updates to the Surgical Technique Guide and IFU on 14 March 2023. The IFU has been updated to address a few small details including the direct listing of single use disposable components and the definition of single use being for one level only. The surgical technique guide has improved graphics, more thorough descriptions for some of the instrument assembly steps that your scrub techs may find useful and a statement regarding the appropriate use of the Flexible Curette Blade to prevent surgeons from using it in a way that may overstress the blade and cause a break. This information was shared with the surgeons through the sole distributor.
Initiated
2023-03-14
Root cause
Device Design
Distribution
US Nationwide distribution in the state of Minnesota.
Recall number
Z-1698-2024
Recall status
Completed
CDRH record ID
206822
FDA classified
2024-04-29
Recall event number
94279

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗