Open original FDA record ↗Recall record
- FDA identifier
- Z-1699-2012
- Company in FDA record
- Ortho-Clinical Diagnostics
- FDA classification date
- Jun 1, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Ortho Clinical Diagnostics is recalling VITROS 5600 Integrated System Software Version 1.6 and below because of an anomaly that may inadvertently add a surfactant to a CurveTip position.
Source facts
- Action
- Ortho Clinical Diagnostics (OCD) sent an Urgent Product Correction Notification letter dated April 26, 2012, via FedEx overnight courier to US consignees with VITROS 5600 Systems. The letter identified the affected product, problem and actions to be taken. Foreign affiliates were informed by email on April 26, 2012 of the issue and instructed to notify their consignees of the issue and actions. OCD recommend customers to group all DAT assays into a single run as described in the letter, until the next Software Version 2.0 is installed on their VITROS 5600 System. Customers were instructed to complete and return the Confirmation of Receipt form. For questions contact Customer Technical Services at 1-800-421-3311
- Initiated
- 2012-04-26
- Root cause
- Software design
- Terminated
- 2018-07-05
- Distribution
- Worldwide Distribution - US (nationwide)
- Recall number
- Z-1699-2012
- Recall status
- Terminated
- CDRH record ID
- 109085
- FDA classified
- 2012-06-01
- Recall event number
- 61790