Skip to content

Public FDA record

Z-1699-2021: NexSite HD, Hemodialysis Symmetric Tip Catheter for long term use

Recall record Z-1699-2021 for MARVAO MEDICAL DEVICES LTD.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1699-2021
Company in FDA record
MARVAO MEDICAL DEVICES LTD.
FDA record creation date
Jun 1, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The action is to ensure consistency between the Directions for Use (DFU) in the field and the DFU on the website.

Source facts

Action
On 02/10/2020, the firm emailed a "Direction For Use Update" Letter, to inform them that the firm has updated its Directions for Use (LS 122) from revision 01 to revision 02 and stated that the update is to provide clarification of an existing warning following a complaint where the stylet hub detached following removal without a guidewire in place. The updated DFU now includes the following statement, "The stylet is not to be removed without the guidewire in place. To remove the stylet forcibly in such cases could result in device damage". Customer are begin informed that the Recalling Firm will be removing LS 122 revision 01 which they currently have and replace it with LS 122 revision 02. This "rework" (replacement) will be conducted by the Recalling Firm's distributor DirectACCESS. The rework (replacement) will occur by the end of March 2020. A copy of the DFU is provided on the Recalling Firm's website and is also attached to the customer notification email. For questions associated with this correction - contact Recalling Firm at: Telephone number - +353 (0) 91 759 301 email - chris.davey@marvaomedical.com or, DirectACCESS at (855)210-0554 Ext 114
Initiated
2020-02-11
Root cause
Labeling Change Control
Distribution
U.S.: CO, NC, SC, TX, O.U.S.: Not provided
Recall number
Z-1699-2021
Recall status
Open, Classified
CDRH record ID
187129
FDA record created
2021-06-01
Recall event number
87827

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗