Open original FDA record ↗Recall record
- FDA identifier
- Z-1699-2024
- Company in FDA record
- Olympus Corporation of the Americas
- FDA classification date
- Apr 30, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Upon initial activation of the seal function, the POWERSEAL non-conformance results in an immediate Incomplete Seal Cycle tone with the accompanying message on the generator screen, and no energy would be delivered to the device resulting in delay in surgery.
Source facts
- Action
- Olympus issued Urgent Medical Device Removal Letter on April 2, 2024 via email/visit. Letter states reason for recall, health risk and action to take: 1. Inspect your inventory and identify any POWERSEAL products having the lot number identified above. Please check all areas of the hospital to determine if any of these devices remain in inventory. Quarantine and cease use of the affected model/lots. 2. Please contact Customer Service at at 1-800-848-9024, option 2, with the quantity, model and lot number of the affected device(s). Olympus will issue a Return Material Authorization to return any affected product at no charge. Olympus will issue a credit to your facility upon return of the affected product. 3. Access the Olympus recall portal to indicate that you have received this notification. a. Go to https://olympusamerica.com/recall b. Enter the recall number "0444 . c. Complete the form as instructed and include your account ID number. 4. If you have distributed these devices outside your facility, please provide a copy of this letter to those facilities immediately. If you require additional information, please do not hesitate to contact me directly by phone at (647) 999-3203 or by e-mail at Cynthia.Ow@olympus.com
- Initiated
- 2024-03-07
- Root cause
- Process control
- Distribution
- Worldwide distribution - US Nationwide and the countries of AU, CA, HK, KR.
- Recall number
- Z-1699-2024
- Recall status
- Open, Classified
- CDRH record ID
- 207122
- FDA classified
- 2024-04-30
- Recall event number
- 94375