Open original FDA record ↗Recall record
- FDA identifier
- Z-1700-2010
- Company in FDA record
- Circaid Medical Products Inc
- FDA classification date
- May 27, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Some systems in these lots have Built-In Pressure System lines with the incorrect amount of spacing. Some systems in these lots had I inch space between each line with no tension applied. The system should have 1.25 inches of space. The correct spacing (1.25") is necessary to achieve the measured tension ranges. The incorrect spacing may result in the inability to achieve the correct tension level
Source facts
- Action
- CircAid Medical Products Important Device Product Recall letter dated Apr 20, 2010 was mailed to consignees describing problem, listing affected product and distribution dates. Customers were instructed to locate and return these Juxta-Lite Systems from stock and notify their customers A Retum Authorization number and contact name was provided. this listing for product returns.
- Initiated
- 2010-04-20
- Root cause
- Process control
- Terminated
- 2011-01-06
- Distribution
- Worldwide distribution: USA, Japan, Canada, Netherlands
- Recall number
- Z-1700-2010
- Recall status
- Terminated
- CDRH record ID
- 91065
- FDA classified
- 2010-05-27
- Recall event number
- 55529