Open original FDA record ↗Recall record
- FDA identifier
- Z-1700-2014
- Company in FDA record
- Hologic, Inc.
- FDA classification date
- Jun 3, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Selenia Full Field Digital Mammography System defective relay switch, if there is a need to use the emergency off switch, the system may not turn off which could result in patient injury
Source facts
- Action
- Hologic notified customers by letter dated 5/13/14 via certified mail or Federal Express to inform customers of the problem and provide acknowledgement of the notification and confirmation that their systems have been corrected A Dealer notification dated 5/13/14 sent by certified mail or Federal Express to inform Dealers of the need to notify their customers of the problem and to correct the identified Selenia system using Hologic Technical Bulletin TB-00806. On 5/16/14 the firm expanded the recall to an additional units Mammography System M-IV (2 units) and Selenia Mammography Systems (4) by letter dated 5/14/14. If there are any questions regarding this field correction, please call Hologic Customer Support at 877 371-4372.
- Initiated
- 2014-05-13
- Root cause
- Component change control
- Terminated
- 2014-08-18
- Distribution
- Distributed in Puerto Rico, and the states of CA, NY, and TX, and the country of Republic of Korea.
- Recall number
- Z-1700-2014
- Recall status
- Terminated
- CDRH record ID
- 127292
- FDA classified
- 2014-06-03
- Recall event number
- 68298