Skip to content

Public FDA record

Z-1702-2015: The Alcon Custom Pak. The Alcon Customer Pak is an assemblage of single-use medical devices and accesso

Recall record Z-1702-2015 for Alcon Research, Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1702-2015
Company in FDA record
Alcon Research, Ltd.
FDA classification date
Jun 9, 2015
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The supplier of the Devon Light Glove, a component of the Alcon Custom PAK, recalled the gloves due to the potential for splits and holes on the glove. Should the user be unaware that the Light Glove is torn/split, surgical site infections are possible.

Source facts

Action
Alcon sent a Recall letter dated April 28, 2015 to affected customers. The letter identified the affected product, problem and actions to be taken. The letter included a return form and instructions for credit. Customers were instructed to review inventory for any affected units, segregat the product to ensure it is not used, and return the attached Response Form via fax or email to Alcon. For questions or need assistance with obtaining replacement products contact Alson Customer service at 800-862-5266.
Initiated
2015-04-28
Root cause
Nonconforming Material/Component
Terminated
2016-03-24
Distribution
Worldwide Distribution - US Nationwide and the countries of Australia and Canada.
Recall number
Z-1702-2015
Recall status
Terminated
CDRH record ID
136291
FDA classified
2015-06-09
Recall event number
71156

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗