Recall record Z-1702-2016 for Angiodynamics Inc. (Navilyst Medical Inc.). View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1702-2016
Company in FDA record
Angiodynamics Inc. (Navilyst Medical Inc.)
FDA classification date
May 20, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Maquet Cardiopulmonary, the manufacturer of the RotaFlow Centrifugal Pump, determined that the products may possess levels of endotoxin that exceed regulatory limits. AngioDynamics confirmed that the affected RotaFlow Centrifugal Pumps were included in packaged AngioVac Circuit Packs.

Source facts

Action
On 3/21/2016, AngioDynamics sent recall notification packages (dated 3/21/2016) to the consignees via Federal Express. Consignees are instructed to follow the instructions provided on the Maquet Cardiopulmonary recall letter (dated 2/23/2016) and return the Reply Verification Tracking Form, provided in the recall notification.
Initiated
2016-03-21
Root cause
Environmental control
Terminated
2017-09-18
Distribution
Nationwide Distribution.
Recall number
Z-1702-2016
Recall status
Terminated
CDRH record ID
145039
FDA classified
2016-05-20
Recall event number
73815

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 21, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 21, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 21, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗