Skip to content

Public FDA record

Z-1702-2018: BARD BrachySource Iodine125 radioactive seeds consist of a welded titanium capsule containing the I125 ab

Recall record Z-1702-2018 for Bard Brachytherapy, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1702-2018
Company in FDA record
Bard Brachytherapy, Inc.
FDA classification date
Apr 27, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Bard Medical Division is recalling the BARD BrachySource Iodine125 Radioactive Seeds in Mick Applicators and BARD BrachySource Iodine125 Radioactive Seeds Loose in Vial with Sterile Spacers. Because the incorrect dosage level radioactive seeds were mixed up and sent to two hospitals. One Hospital ordered BrachySource Iodine125 0.36 mCi radioactive seeds but instead received BrachySource Iodine125 0.28 mCi radioactive seeds and vice versa for the other hospital affected.

Source facts

Action
The notification to the customer was completed by Bard Medical Customer Service via a telephone call to alert them of the incorrect product shipped on April 10, 2018. Bard Medical Division is recalling the BARD BrachySource Iodine125 Radioactive Seeds in Mick Applicators and BARD BrachySource Iodine125 Radioactive Seeds Loose in Vial with Sterile Spacers. The incorrect dosage level of radioactive seeds was sent to 2 hospitals. One hospital ordered 0.36 mCi radioactive seeds but instead received 0.28 mCi radioactive seeds intended for the other hospital. For further questions, please call (770) 385-0562.
Initiated
2018-04-10
Root cause
Labeling mix-ups
Terminated
2020-06-05
Distribution
US Distribution to the states of : California and Georgia
Recall number
Z-1702-2018
Recall status
Terminated
CDRH record ID
164063
FDA classified
2018-04-27
Recall event number
79972

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗