Public FDA record

Z-1702-2024: Abbott Cholestech LDX Battery Kit, Model: 3LP656194-PCM-LD, PN: 55170

Recall record Z-1702-2024 for Alere San Diego, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1702-2024
Company in FDA record
Alere San Diego, Inc.
FDA classification date
Apr 30, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a potential that the Cholestech LDX battery (an external battery pack available as an optional accessory to the Cholestech LDX Analyzer) may swell which could cause minor burns, cuts or abrasions, or low current electrical shock to the user.

Source facts

Action
On March 28, 2024, Alere San Diego issued a Urgent Medical Device Correction notification to affected consignees via mail. In addition to informing consignees about the recall, Alere asked consignees to take the following actions: " Inspect all Cholestech LDX" Battery Kits for any signs of swelling. DO NOT CONTINUE USE OF BATTERIES THAT APPEAR SWOLLEN. If your battery pack exhibits any signs of swelling, the product should not be used and discarded per local regulations. " If the battery appears normal, then it can continue to be used. " Complete and return the Customer Reply Form. " If you have forwarded the products listed above to other laboratories, please inform them of this Product Correction and provide to them a copy of this letter. " Please retain this letter for your laboratory records.
Initiated
2024-03-01
Root cause
Labeling design
Distribution
Worldwide distribution - US Nationwide and the countries of United Kingdom, Netherlands and Kenya.
Recall number
Z-1702-2024
Recall status
Open, Classified
CDRH record ID
206539
FDA classified
2024-04-30
Recall event number
94190

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗