Open original FDA record ↗Recall record
- FDA identifier
- Z-1702-2024
- Company in FDA record
- Alere San Diego, Inc.
- FDA classification date
- Apr 30, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
There is a potential that the Cholestech LDX battery (an external battery pack available as an optional accessory to the Cholestech LDX Analyzer) may swell which could cause minor burns, cuts or abrasions, or low current electrical shock to the user.
Source facts
- Action
- On March 28, 2024, Alere San Diego issued a Urgent Medical Device Correction notification to affected consignees via mail. In addition to informing consignees about the recall, Alere asked consignees to take the following actions: " Inspect all Cholestech LDX" Battery Kits for any signs of swelling. DO NOT CONTINUE USE OF BATTERIES THAT APPEAR SWOLLEN. If your battery pack exhibits any signs of swelling, the product should not be used and discarded per local regulations. " If the battery appears normal, then it can continue to be used. " Complete and return the Customer Reply Form. " If you have forwarded the products listed above to other laboratories, please inform them of this Product Correction and provide to them a copy of this letter. " Please retain this letter for your laboratory records.
- Initiated
- 2024-03-01
- Root cause
- Labeling design
- Distribution
- Worldwide distribution - US Nationwide and the countries of United Kingdom, Netherlands and Kenya.
- Recall number
- Z-1702-2024
- Recall status
- Open, Classified
- CDRH record ID
- 206539
- FDA classified
- 2024-04-30
- Recall event number