Open original FDA record ↗Recall record
- FDA identifier
- Z-1702-2026
- Company in FDA record
- Synthes (USA) Products LLC
- FDA classification date
- Mar 30, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
A full lot swap between VOLT Locking Screw Tap 03.424.124 lot 68900P9 and Cortical Screw Tap 03.424.126 lot 68982P8, resulting in the bones taps having an incorrect thread.
Source facts
- Action
- Depuy Synthes initially notified Sales Consultants and Regional Managers on about 02/25/2026, followed by notification of consignees. Consignees were instructed to examine inventory for affected units, quarantine them immediately, and contact Depuy Synthes sales consultant or customer support services to arrange the return/credits for affected units. Additionally, consignees were requested to complete and return the provided Business Response Form as well as notify all applicable personnel within the facility or any consignees if the affected units were further distributed. Consignees were also requested to post the notification in a visible area for awareness.
- Initiated
- 2026-02-25
- Root cause
- Process change control
- Distribution
- US Nationwide distribution in the states of AR, AZ, CA, CO, FL, IN, LA, MD, MN, NC, NJ, OH, OK, OR, PA, TN, TX, UT, VA, WA, WI.
- Recall number
- Z-1702-2026
- Recall status
- Open, Classified
- CDRH record ID
- 218813
- FDA classified
- 2026-03-30
- Recall event number
- 98461